GRAPHITE 52M Heated FID Total Hydrocarbon (THC, COV, CH4) Analyzer

GRAPHITE 52M Heated FID Total Hydrocarbon (THC, COV, CH4) Analyzer

3 versions available :

  • GRAPHITE 52M-S : Total Hydrocarbon monitoring (THC, COV)
  • GRAPHITE 52M-D : Total Hydrocarbon (THC, COV) and Methane (CH4) simultaneous monitoring
  • GRAPHITE 52M-W : Total Hydrocarbon (THC, COV) simultaneous monitoring on 2 independent streams

Applications:

  • Engine Exhaust gas analysis
  • Standard Reference Method for Emission monitoring
  • Laboratories and research centres
  • Process control in all fields
  • Combustion monitoring in all fields

Features:

  • Burners placed in a heated furnace, up to 191°C allowing measurement of high concentrations of heavy hydrocarbons
  • Ultra short response time
  • Graphic Liquid Crystal Display (LCD)
  • Interactive menu driven software with enhanced speed display
  • Real time calibration graph
  • User adjustable response time and averaging time
  • Long-life catalysator with high efficiency
  • Built-in storage of 2 months 1/4 h average data (up to 1 year with the optional memory extension)
  • Internal zero and burner air scrubber
  • Built-in double serial interface (RS 232/RS 422) for remote control & maintenance
  • Ethernet network connection

Specifications:

  • Ranges: 0-10/100/1 000/10 000 ppm or user programmable range up to 50 000 ppm
  • Accuracy: < 1% of the displayed value between 15% and 100% of the F.S.
  • Lower detectable limit: 0.05 ppm on the 10 ppm range
  • Response time: < 1 sec. (T90)
  • Zero drift: < 1% / 24 h F.S. < 2% / 7 days F.S.
  • Span drift: < 1% / 7 days F.S.
  • Linearity: < 1 % for concentrations between 10 % and 90% of the scale
  • Temperature of the heated block: up to 191°C

Certifications:

Certified TÜV and MCERTs:

  • United Kingdom : Sira, MCERT n° MC 060082/01
  • Germany: TÜV, ID 0000035009 : Compliant with EN 15267-1:2009 , EN 15267-2:2009, EN 15267-3:2007, EN 14181 : 2004
  • Complies with CFR 40-1065 for engine exhaus gas monitoring

more..