MIR 9000 – MULTI-GAS ANALYZER INFRA-RED GFC

MIR 9000 – MULTI-GAS ANALYZER INFRA-RED GFC 

A unique multi-gaz analyzer for: NO, NO2, NOx, N2O, CO, CO2, SO2, HCl, HF, CH4, TOC and O2. Optional measurement of NOX by chemiluminescence.

  • Over 3.000 MIR 9000 multi-gas analyzer based installations worldwide, in a wide range of applications and industries
  • External data acquisition of up to 7 additional parameters (flow, pressure, temperature or other analog input)
  • Chemiluminescence module optionally (CLD) for low NO, NOx and NO2 measurement
  • Built-in paramagnetic cell for O2 monitoring
  • Real time graphic display
  • Interactive menu-driven software allowing ease of operation
  • Unheated sample line utilising the permeation drying technology
  • Readings measured and expressed on a dry basis
  • On-board oxygen correction for environmental reporting
  • Automatic cross interference correction
  • Highly accurate, excellent stability with automatic optical stability check
  • Intrinsic security with on-board residual H2O measurement
  • Available in 19’’ Rack or Tight box version

- MCERTs and TÜV certified to EN15267-3

- QAL1 as defined by EN14181

- QAL3 compliance to EN14181

- Exceeds U.S EPA 40 CFR 60 and 75 performance specifications

Аpplications:

Municipal and Hazardous Waste Incinerators

  • Cogeneration, Gas Turbines (MIR 9000 with CLD option)
  • Industrial Boilers and Furnaces
  • Process monitoring
  • Power & Combustion
  • Cement Kilns, Glass, Chemical & Petrochemical Plants

Features:

Infra-Red Gas Filter Correlation allows the MIR 9000 to offer up to ten gas measurements simultaneously, including Nitric Oxide (NO), Nitrogen Dioxide (NO2), Sulphur Dioxide (SO2), Carbon Monoxide (CO), even Hydrogen Chloride (HCl) and Hydrogen Fluoride (HF).

Complete systems would normally comprise of:

  • Sample extraction and conditioning probe (with integrated temperature, pressure and flow measurement)
  • Sample lines
  • Automatic calibration units
  • Instrument air drying system
  • Data acquisition & management system.

Specifications:

  • Repeatability: <2% of Full Scale (F.S.)
  • Zero drift: <2% F.S. / 30 days
  • Span drift: < 1% F.S. / 7 days
  • Linearity: < 1% F.S.
  • Power supply: 80 - 230V, 50-60 Hz
  • Consumption: 300 VA
  • Serial link: RS232, RS 422
  • Operating temperature: +5 °C to +40 °C

Version with CLD (tight box):

  • Dim.: 200x600x600 mm (DxWxH)
  • Weight: 32 Kg

Version without CLD (tight box):

  • Dim.: 200x400x600 mm (DxWxH)
  • Weight: 24 Kg

Version without CLD (19’’ Rack):

  • Dim.: 490x483x177 mm (DxWxH)
  • Weight: 14 Kg

MAIN OPTIONS:

  • Pressure, temperature & gas velocity measurements
  • SEC® sampling system (permeation based)
  • Analog output: 0/10V - 0/4-20 mA programmable
  • Rack cabinet, cubicle or shelter integration
  • CONTACT™ software for remote maintenance
  • O2, measurement via built-in paramagnetic cell
  • TIG calibration / back flush module
  • MDS dryer
  • Multiplexing system
  • Data acquisition and management system WEX®
  • CLD module for low concentrations of NO, NOx and NO2

Certifications:

MCERTs and TÜV certified to EN15267-3

  • QAL1 as defined by EN14181
  • QAL3 compliance to EN14181
  • Exceeds U.S EPA 40 CFR 60 and 75 performance specifications 

more..