MIR IS
MIR IS
The unique AMS on CEMS market offering multi gas accurate multi-pollutant analysis integrating optionally flue gas temperature, flow and pressure measurements on a single sampling probe, means a single sampling point.
- Direct installation on stack or duct.
- Designed to operate under legislation such as 2000/76/EC (WID) and 2001/80/EC (LCPD),
- The MIR-IS offers maximum availability and complete compliance with QAL1 and QAL3 of EN14181 & EN15267-3.
Аpplications:
- Combustion plants
- Municipal waste incinerators (upstream measurements)
- Hazardous waste incinerators (industrial and hospital)
- Gas treatment process control
- Paper mills
- Chemical industries
Features:
- In situ short extractive analyzer, with built-in sample treatment module
- Short response time
- Measured gases: from 1 up to 11
- Acquisition and processing of 5 additional measurements (flow rate, pressure, gas temperature...) or any other analog input
- 2 processors (1 is dedicated to communication and remote maintenance processing)
- Graphic LCD screen with real time synoptic
- Pop up menus, with graphic functions
- Powerful remote control capabilities
- Measurement ranges: consult the pdf brochure
- Repeatability: ± 2% of full scale (F.S.)
- Zero drift: ± 2% F.S. / 30 days
- Span drift: ± 2% F.S. / 30 days
- Linearity: ± 1% F.S.
- Power supply: 220 V, 50-60 Hz, 700 VA
- Housing: IP 65 tight box
- Dimensions: 200 x 600 x 600 m (DxWxH)
- Weight: 35 kg approx.
- Operating temperature: - 15°C to + 50°C
- Instrument air : 7Nm3/h, 5 bars
Certifications: