» Products» Газ анализатори» AMESA - Continuous dioxin/furan emission monitoring by long-term sampling
AMESA - Continuous dioxin/furan emission monitoring by long-term sampling
AMESA - Continuous dioxin/furan emission monitoring by long-term sampling
The only long-term sampling system for dioxins with TÜV, MCERT and ETV EPA approvals
Applications:
- Municipal waste incinerators
- Hazardous waste incinerators
- Hospital waste incinerators
- Power plants
- Biomass-fired power plants
- Cement kilns
- Metallurgic plants
- Pulp and paper
Features:
- Modified cooled-probe method, with adsorption on XAD-II
- Collection of condensate not required
- TÜV approved and type performance tested by TÜV Rheinland (Germany)
- Certified by MCERT (UK)
- Fully automatic isokinetic sampling from 6 hours up to 6 weeks (4 weeks type performance tested)
- Validated against method EN 1948
- US EPA method 23A compliant (option)
- Suitable for all U-POPs listed in the Stockholm convention
- Optional upgrade for sampling of other micro-pollutants (e.g. heavy metals or mercury)
Specifications:
- Measuring range (dioxin & furan): .0.0001-10 ng I-TEQ / m3 (TÜV approved 0-0.2 ng I-TEQ / m3)
- Sampling interval: from 6 hours up to 6 weeks (TÜV approved 4 weeks)
- Flue gas temperature: up to 70°C without cooling (up to 400°C with cooling)
- Max dust concentration in the flue gas: 50 mg/ m3
- Flue gas velocity:.from 1 up to 30 m/s
- Operating temperature (control cabinet): 5 to +40°C (optional air conditioner for temperatures over +40°C)
- Max relative humidity (control cabinet): 50%
- Isokinetic control cycle: 1 sec
- Velocity measurement accuracy: ± 1% of measuring range
- Volume measurement accuracy: ± 1.5% of measuring range
Certifications:
- Germany : TUVdotCom-ID: 0011005400, certificate n° 936/805017B)
- United Kingdom : MCERTS (n° Sira MC 050064/00)