TOPAZE 32M
TOPAZE 32M
Utilizes the CLD technique for continuous and simultaneous measurement of NO, NO2 and NOx, in compliance with EN14792.
Exists in 2 different versions, to meet your analytical requirements:
TOPAZE 32M-S: NO or NOx monitoring
TOPAZE 32M-D: NO-NOx and NO2 simultaneous monitoring
Аpplications:
- Engine manufacturing and testing in the automotive, heavy duty, aviation, marine, rail transport sectors, etc.
- Continuous Emission Monitoring
- Combustion control
- Process monitoring
- Laboratories and Research Centres...
Features:
- Proven EPA reference method
- Accurate measurement in the range: 0-10/100/1 000/10 000 ppm
- Heated analyzer (temperature controlled up to 180°C), measuring chamber under vacuum resulting minimized quenching effect
- Graphic Liquid Crystal Display (LCD)
- Interactive menu driven software with enhanced speed display
- Real time calibration graph
- Built-in memory for data storage
- Built-in Ethernet TCP/IP connection for remote control and serial interface RS 232
- Fast response time
- Communication by AK protocol (RS232 / RJ45)
Specifications:
- Ranges: 0-10/100/1 000/10 000 ppm
- Noise (2 sigma): < 1% F.S.
- Lower detection limit: 0.1 ppm on the lowest range
- Response time: < 2 sec (T 10-90)
- Zero drift: ± 1% / 24 h
- Span drift: ± 1 % / 24 h
- Linearity: ± 1 % for a concentration between 15 % and 100% of the full scale’s range
- Sample flow rate: 0.7 to 1 l/min at 20 psi
- Capillary block temperature: heated up to 180°C
- Housing: Standard 19 ‘’ – 3U rack
- Dimensions: 483 x 440 x 135 mm (L x W x H) / 19 x 17.3 x 5.3 inches (L x W x H)
- Weight: 10 kg / 22 lbs
- Operating temperature: 5 - 45 °C
- Communication: RS 232 & TCP/IP, AK protocol
- Ethernet network connection